LPC-310 LCD (1CFM)

2014-04-10 11:39 PV: 3066  Hot

LPC-310 laser particle counter is used to measure the size and quantity of particles within air of unit volume in clean environment; it can directly measure clean environments from clean class 100,000 to 10. This counter has used semiconductor laser power and nixie tube with high-luminance, so it is small, light, highly accurate, easing-operating and microprocessor-controlled. It may store and print the measure results, the printed result will show the clean class of that point and the measure of the clean environment is convenient.

The product is widely used in electronics, optics, chemistry, food, cosmetic, medical science, biological products, aviation, and so on.

LPC-310 Laser Particle Counter of Big Charge LCD

Characteristics:


high accuracy(1CFM)
thermal printer inside, can be transferred to M3
small, light ,easy-operating, convenient
short self-purifying time
with flowmeter itself

Main Technical Parameter:
  

Model

LPC-310 LCD

LPC-310 LCD Pile

Dimensions

350×420×142mm3 (W×D×H)

Display

LCD

Weight

10kg

Maximum power dissipation

80W

60W

Power supply

AC100-220V  50/60Hz

AC(100-240V 50/60Hz) and 
DC(14.8V 14000mA Li-ion batteries

Particle size passage

0.30.51.03.05.010 μm

Sampling flow rate

1CFM(28.3L/min)

Measure period

6s/30s/1-40min

Self-purifying time

≤10min

print

Thermal printer inside, can be printed into grain/m3

Warning class

100100010000

Usable environmental condition

Temperature 1030

   

Humidity: 20%75%

   

Atmospheric pressure86Pa-106kPa

Maximum sampling concentration permitted

35000 grain/Lthe particle size of a particle shall not be over 0.5μm

acid ,alkali and such corrosive gas are not permitted in the sampling air.

Continuous work time

8 hours

About 4 hours

Light source and life

Semiconductor laserlife is over 30,000 hours

Packing of data

256 groups


 

 

 